Skip to Main content Skip to Navigation
Journal articles

Multidimensional Calibration of the EXOGAM Segmented Clover

Abstract : The EXOGAM clover detector consists of four high-purity germanium (HPGe) crystals in a common cryostat. Each individual crystal is electrically segmented into four regions. We propose a method based on a multidimensional fitting procedure to calibrate the energy from the segments. This method could be used for any kind of segmented germanium detector. Its application to the EXOGAM clover is discussed in details.
Document type :
Journal articles
Complete list of metadata

http://hal.in2p3.fr/in2p3-00112054
Contributor : Sylvie Flores <>
Submitted on : Tuesday, November 7, 2006 - 11:27:45 AM
Last modification on : Thursday, June 17, 2021 - 3:18:28 PM
Long-term archiving on: : Tuesday, April 6, 2010 - 9:42:19 PM

Identifiers

Collections

Citation

B. Rossé, O. Stezowski. Multidimensional Calibration of the EXOGAM Segmented Clover. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2006, 565, pp.623-629. ⟨10.1016/j.nima.2006.05.211⟩. ⟨in2p3-00112054⟩

Share

Metrics

Record views

595

Files downloads

587