Spectral discrimination of Cerenkov radiation in scintillating dosimeters - IN2P3 - Institut national de physique nucléaire et de physique des particules Accéder directement au contenu
Article Dans Une Revue Medical Physics Année : 2005

Spectral discrimination of Cerenkov radiation in scintillating dosimeters

Résumé

Radiation therapy accelerators require highly accurate dose deposition and the output must be monitored frequently and regularly. Ionization chambers are the primary tool for this control, but their size, their high voltage needed, and the correction needed for electrons make them unsuitable for use during patient treatment. We have developed a small (1-mm-diam and 1-mm-long active part), flexible, and water-equivalent dosimeter. It is suitable for photon and electron beams without corrections, and performs on line dose measurements. This detector is based on only one scintillating fiber and a CCD camera. A new signal processing is used to remove the effect of Cerenkov radiation background, which only requires a preliminary calibration. Central-axis depth-dose distribution comparisons have been achieved with standard ionization chambers, over a range from 8 to 25 MV photons and from 6 to 21 MeV electrons in order to validate this calibration. Results show a very good agreement, with less than 1% difference between the two detectors.
Fichier non déposé

Dates et versions

in2p3-00113799 , version 1 (21-11-2006)

Identifiants

Citer

A.M. Frelin-Labalme, J.M. Fontbonne, G. Ban, J. Colin, M. Labalme, et al.. Spectral discrimination of Cerenkov radiation in scintillating dosimeters. Medical Physics, 2005, 32, pp.3000-3006. ⟨10.1118/1.2008487⟩. ⟨in2p3-00113799⟩
31 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More