On-line monitoring of fluence distributions and imaging of scanning ion beams

Abstract : In this paper we describe the design of an ion beam monitor developed to control irradiations of biological samples with ions at GANIL (Grand Accélérateur National d'Ions Lourds). This device can be used as an on-line monitor to provide transverse fluence distributions for active scanning ion irradiations. It can also be used as an interceptive beam imager to provide beam profiles or fluence depth distributions. A prototype of the monitor has been tested at GANIL with five different ion beams, having intensities from 104 to 109ions per second. Real time transverse fluence distributions have been obtained with a 1mm spatial resolution.
Document type :
Journal articles
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00258718
Contributor : Sandrine Guesnon <>
Submitted on : Monday, February 25, 2008 - 9:36:24 AM
Last modification on : Tuesday, February 19, 2019 - 9:56:08 AM
Long-term archiving on : Thursday, May 20, 2010 - 6:42:01 PM

File

NIM_A_Pautard_2007.pdf
Files produced by the author(s)

Identifiers

Citation

C. Pautard, E. Balanzat, G. Ban, E. Batin, B. Carniol, et al.. On-line monitoring of fluence distributions and imaging of scanning ion beams. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2008, 588, pp.448-456. ⟨10.1016/j.nima.2008.01.098⟩. ⟨in2p3-00258718⟩