Skip to Main content Skip to Navigation
Journal articles

A fast online hit verification method for the singl ion hit system at GSI

Abstract : For a single ion hit facility built to irradiate specific targets inside biological cells, it is necessary to prove that the ions hit the selected targets reliably because the ion hits usually cannot be seen. That ability is traditionally tested either indirectly by aiming at pre-etched tracks in a nuclear track detector or directly by making the ion tracks inside cells visible using a stain coupled to special proteins produced in response to ion hits. However, both methods are time consuming and hits can be verified only after the experiment. This means that targeting errors in the experiment cannot be corrected during the experiment. Therefore, we have developed a fast online hit verification method that measures the targeting accuracy electronically with a spatial resolution of ±1 µm before cell irradiation takes place.
Document type :
Journal articles
Complete list of metadata

http://hal.in2p3.fr/in2p3-00292860
Contributor : Pascale Chambon <>
Submitted on : Thursday, July 3, 2008 - 9:38:18 AM
Last modification on : Thursday, January 11, 2018 - 6:12:51 AM

Links full text

Identifiers

Collections

Citation

G. Du, B. Fischer, P. Barberet, M. Heiss. A fast online hit verification method for the singl ion hit system at GSI. Radiation Protection Dosimetry, Oxford University Press (OUP), 2006, 122, pp.320-322. ⟨10.1093/rpd/ncl435⟩. ⟨in2p3-00292860⟩