Skip to Main content Skip to Navigation
Poster communications

A New Intraoperative Gamma Camera for the Sentinel Lymph Node Procedure in Breast Cancer

Abstract : This study aimed at evaluating the performance of an intraoperative gamma camera, named CarolIReS, to detect axillary drainage and to assess the removal of sentinel lymph nodes (SLN) in breast surgery. Patients and Methods: SLN biopsy was performed on 25 patients and the CarolIReS camera was used preoperatively to localize SLNs. During surgery, individual removal of SLNs was performed using a gamma probe and their activity was measured with a gamma ray counter. At the end of surgery, the CarolIReS camera was used again to check the quality of surgery which was followed by surgical excision for remaining SLNs. Results: The detection efficiency of the CarolIReS camera was 2.2 cps/kBq for 99mTc activity in SLNs. In one case, it allowed the detection of a residual SLN with a low activity (0.5 kBq) which was massively metastatic. Conclusion: Intraoperative cameras could be used to improve the efficiency of the SLN procedure
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00368761
Contributor : Yvette Heyd <>
Submitted on : Tuesday, March 17, 2009 - 3:27:52 PM
Last modification on : Friday, May 8, 2020 - 1:10:19 AM

Identifiers

  • HAL Id : in2p3-00368761, version 1

Collections

Citation

C. Mathelin, S. Salvador, V. Bekaert, S. Croce, N. Andriamisandratsoa, et al.. A New Intraoperative Gamma Camera for the Sentinel Lymph Node Procedure in Breast Cancer. 8th International Conference of Anticancer Research, Oct 2008, Kos, Greece. 28, pp.2859-2864, 2008. ⟨in2p3-00368761⟩

Share

Metrics

Record views

108