Measurement of the Non-Linearity and Interpixel Capacitance of a H2RG (2Kx2K) Near-IR Detector

Abstract : A temporal analysis of the noise of a H2RG detector is performed in view of the calibration of a dedicated setup. The implication on the noise performance of the subtraction of the reference output is described, and we extend the correlation method to groups of several pixels to derive the ratio x of the interpixel capacitance to the pixel capacitance of a detector. We find View the MathML source. All measurements are consistent to a sub-percent accuracy and provide a calibration of our setup.
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00467907
Contributor : Danielle Cristofol <>
Submitted on : Monday, March 29, 2010 - 3:19:55 PM
Last modification on : Tuesday, November 19, 2019 - 2:38:30 AM

Identifiers

Collections

Citation

G. Smadja, C. Cerna, A. Ealet. Measurement of the Non-Linearity and Interpixel Capacitance of a H2RG (2Kx2K) Near-IR Detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2009, 610, pp.615-621. ⟨10.1016/j.nima.2009.08.070⟩. ⟨in2p3-00467907⟩