Skip to Main content Skip to Navigation
Journal articles

Planck pre-launch status: The HFI instrument, from specification to actual performance

J.-M. Lamarre 1 J.-L. Puget 2 P.A.R. Ade F. Bouchet G. Guyot 2 A. E. Lange F. Pajot 2 A. Arondel 2 K. Benabed 3 J.-L. Beney 4 Alain Benoit 5 J.-Ph. Bernard 6 R. Bhatia Y. Blanc J. J. Bock E. Bréelle 7 T. W. Bradshaw Philippe Camus 8 A. Catalano 7 J. Charra 2 M. Charra 2 S. E. Church F. Couchot 4 A. Coulais 1 B. P. Crill M. R. Crook K. Dassas 2 P. de Bernardis J. Delabrouille 9 P. de Marcillac 2 J.-M. Delouis 3 F.-X. Désert 10 C. Dumesnil X. Dupac 6 G. Efstathiou P. Eng 2 C. Evesque 2 J.-J. Fourmond 2 K. Ganga 11 M. Giard 6 R. Gispert 2 L. Guglielmi 7 J. Haissinski 4 S. Henrot-Versillé 4 E. Hivon 3 W. A. Holmes W. C. Jones T. C. Koch H. Lagardère P. Lami 2 J. Landé B. Leriche 2 C. Leroy Y. Longval 2 J.F. Macías-Pérez 12 T. Maciaszek 13 B. Maffei 1 B. Mansoux 4 C. Marty 6 S. Masi C. Mercier 2 M.-A. Miville-Deschênes 2 A. Moneti 3 L. Montier 6 J. A. Murphy J. Narbonne 6 M. Nexon C. G. Paine J. Pahn O. Perdereau 4 F. Piacentini M. Piat 7 Stéphane Plaszczynski 4 E. Pointecouteau 6 R. Pons N. Ponthieu 2 S. Prunet 3 D. Rambaud 6 G. Recouvreur 2 C. Renault 12 I. Ristorcelli 6 C. Rosset 7 D. Santos 12 G. Savini G. Serra 6 P. Stassi 12 R.V. Sudiwala J.F. Sygnet 3 J. A. Tauber J.-P. Torre 2 M. Tristram 4 L. Vibert 2 A. Woodcraft V. Yurchenko D. Yvon 14
Abstract : Context. The High Frequency Instrument (HFI) is one of the two focal instruments of the Planck mission. It will observe the whole sky in six bands in the 100 GHz-1 THz range. Aims: The HFI instrument is designed to measure the cosmic microwave background (CMB) with a sensitivity limited only by fundamental sources: the photon noise of the CMB itself and the residuals left after the removal of foregrounds. The two high frequency bands will provide full maps of the submillimetre sky, featuring mainly extended and point source foregrounds. Systematic effects must be kept at negligible levels or accurately monitored so that the signal can be corrected. This paper describes the HFI design and its characteristics deduced from ground tests and calibration. Methods: The HFI instrumental concept and architecture are feasible only by pushing new techniques to their extreme capabilities, mainly: (i) bolometers working at 100 mK and absorbing the radiation in grids; (ii) a dilution cooler providing 100 mK in microgravity conditions; (iii) a new type of AC biased readout electronics and (iv) optical channels using devices inspired from radio and infrared techniques. Results: The Planck-HFI instrument performance exceeds requirements for sensitivity and control of systematic effects. During ground-based calibration and tests, it was measured at instrument and system levels to be close to or better than the goal specification.
Complete list of metadatas

Cited literature [21 references]  Display  Hide  Download

http://hal.in2p3.fr/in2p3-00535714
Contributor : Dominique Girod <>
Submitted on : Friday, November 12, 2010 - 2:40:10 PM
Last modification on : Friday, September 25, 2020 - 11:16:03 AM
Long-term archiving on: : Sunday, February 13, 2011 - 2:46:27 AM

File

aa12975-09.pdf
Publisher files allowed on an open archive

Identifiers

Citation

J.-M. Lamarre, J.-L. Puget, P.A.R. Ade, F. Bouchet, G. Guyot, et al.. Planck pre-launch status: The HFI instrument, from specification to actual performance. Astronomy and Astrophysics - A&A, EDP Sciences, 2010, 520, pp.A9. ⟨10.1051/0004-6361/200912975⟩. ⟨in2p3-00535714⟩

Share

Metrics

Record views

1646

Files downloads

1398