Measurements of high-energy g-rayswithLaBr3 : Ce detectors

Abstract : The full-energy peak efficiency calibration and the energy resolution measurements of the image LaBr3 γ-ray detector are presented for γ-ray energies in the 700 keV–17.6 MeV range. Measurements were done using a combination of proton-capture nuclear reactions on image, image, image, image and image for high-energy γ-rays, and radioactive sources such as image and image for the lowest energies. At high energies, two γ-rays in a cascade from proton resonance capture were employed using Al, Na2WO4, K2SO4 and LiBO2 targets. The obtained results were compared to the simulations performed using a GEANT4 code.
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00575605
Contributor : Christine Hadrossek <>
Submitted on : Friday, March 11, 2011 - 2:47:38 PM
Last modification on : Wednesday, September 16, 2020 - 4:17:41 PM

Identifiers

Collections

IN2P3 | CEA | CSNSM | CNRS | DAM

Citation

M. Ciemała, D. Balabanski, M. Csatlós, J.M. Daugas, G. Georgiev, et al.. Measurements of high-energy g-rayswithLaBr3 : Ce detectors. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2009, 608, pp.76-79. ⟨10.1016/j.nima.2009.06.019⟩. ⟨in2p3-00575605⟩