Skip to Main content Skip to Navigation
Journal articles

Verifying the accuracy of the TITAN Penning-trap mass spectrometer

M. Brodeur V.L. Ryjkov T. Brunner S. Ettenauer A.T. Gallant V.V. Simon M.J. Smith A. Lapierre R. Ringle P. Delheij M. Good D. Lunney 1 J. Dilling
1 SNO
CSNSM - Centre de Spectrométrie Nucléaire et de Spectrométrie de Masse
Abstract : TITAN (TRIUMF's Ion Traps for Atomic and Nuclear science) is an online facility designed to carry out high-precision mass measurements on singly and highly charged radioactive ions. The TITAN Penning trap has been built and optimized in order to perform such measurements with an accuracy in the sub ppb-range. A detailed characterization of the TITAN Penning trap is presented and a new compensation method is derived and demonstrated, verifying the performance in the range of sub-ppb.
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00644148
Contributor : Christine Hadrossek <>
Submitted on : Friday, November 25, 2011 - 2:15:29 PM
Last modification on : Wednesday, September 16, 2020 - 5:38:29 PM

Links full text

Identifiers

Collections

Citation

M. Brodeur, V.L. Ryjkov, T. Brunner, S. Ettenauer, A.T. Gallant, et al.. Verifying the accuracy of the TITAN Penning-trap mass spectrometer. International Journal of Mass Spectrometry, Elsevier, 2011, 310, pp.20-31. ⟨10.1016/j.ijms.2011.11.002⟩. ⟨in2p3-00644148⟩

Share

Metrics

Record views

112