Self-adjusting resistor unit in the detector-voltage supply of a charge-sensitive pre-amplifier for an ionization detector

K.-H. Schmidt 1 B. Jurado 1
1 ACEN - Aval du cycle et Energie Nucléaire
CENBG - Centre d'Etudes Nucléaires de Bordeaux Gradignan
Abstract : A self-adjusting resistor unit is described which stabilizes the bias voltage of a detector for ionising particles, while always keeping the parallel thermal noise of the bias resistor significantly below the detector shot noise. The user is exempted from adjusting the bias resistor to the detector leakage current for achieving optimum operation conditions. Moreover, the unit avoids the drop of the operation voltage on the detector when the reverse current increases up to several orders of magnitude without any further intervention. The self-adjusting unit leads to an extension of the time of operation of the detector under stable conditions by up to several orders of magnitude.
Type de document :
Article dans une revue
IEEE Transactions on Nuclear Science, Institute of Electrical and Electronics Engineers, 2012, 60, pp.2918-2922. 〈10.1109/TNS.2013.2270013〉
Liste complète des métadonnées

Littérature citée [4 références]  Voir  Masquer  Télécharger

http://hal.in2p3.fr/in2p3-00725852
Contributeur : Nathalie Martin <>
Soumis le : mardi 28 août 2012 - 08:48:02
Dernière modification le : lundi 13 octobre 2014 - 15:43:25
Document(s) archivé(s) le : vendredi 16 décembre 2016 - 08:29:07

Fichier

hal-1.pdf
Fichiers produits par l'(les) auteur(s)

Identifiants

Collections

Citation

K.-H. Schmidt, B. Jurado. Self-adjusting resistor unit in the detector-voltage supply of a charge-sensitive pre-amplifier for an ionization detector. IEEE Transactions on Nuclear Science, Institute of Electrical and Electronics Engineers, 2012, 60, pp.2918-2922. 〈10.1109/TNS.2013.2270013〉. 〈in2p3-00725852〉

Partager

Métriques

Consultations de
la notice

136

Téléchargements du document

243