Skip to Main content Skip to Navigation
Conference papers

High performance detector head for PET and PET/MR with continuous crystals and SiPMs

Abstract : A high resolution PET detector head for small animal PET applications has been developed. The detector is composed of a 12 mm x 12 mm source continuous LYSO crystal coupled to a 64-channel monolithic SiPM matrix from FBK-irst. Crystal thicknesses of 5 mm and 10 mm have been tested, both yielding an intrinsic spatial resolution around 0.7 mm FWHM with a position determination algorithm that can also provide depth-of-interaction information. The detectors have been tested in a rotating system that makes it possible to acquire tomographic data and reconstruct images of 22Na sources. An image reconstruction method specifically adapted for continuous crystals has been employed. The Full Width at Half Maximum measured from a point source reconstructed with ML-EM was 0.7 mm with the 5 mm crystal and 0.8 mm with the 10 mm crystal.
Complete list of metadatas

http://hal.in2p3.fr/in2p3-00730629
Contributor : Sabine Starita <>
Submitted on : Monday, September 10, 2012 - 4:52:25 PM
Last modification on : Wednesday, September 16, 2020 - 5:43:29 PM

Identifiers

Collections

IN2P3 | LAL | CNRS

Citation

G. Llosa, P. Barrillon, J. Barrio, M.G. Bisogni, J. Cabello, et al.. High performance detector head for PET and PET/MR with continuous crystals and SiPMs. PET/MR and SPECT/MR: New Paradigms for Combined Modalities in Molecular Imaging Conference, May 2012, La Biodola, Italy. pp.3-5, ⟨10.1016/j.nima.2012.08.099⟩. ⟨in2p3-00730629⟩

Share

Metrics

Record views

131