Skip to Main content Skip to Navigation
Journal articles

A New Front-End High-Resolution Sampling Board for the New-Generation Electronics of EXOGAM2 and NEDA Detectors

Abstract : This paper presents the final design and results of the FADC Mezzanine for the EXOGAM (EXOtic GAMma array spectrometer) and NEDA (Neutron Detector Array) detectors. The measurements performed include those of studying the effective number of bits, the energy resolution using HP-Ge detectors, as well as timing histograms and {\rm n}/\gamma discrimination performance. Finally, the conclusion shows how a common digitizing device has been integrated in the experimental environment of two very different detectors which combine both low-noise acquisition and fast sampling rates. Not only the integration fulfilled the expected specifications on both systems, but it also showed how a study of synergy between detectors could lead to the reduction of resources and time by applying a common strategy.
Complete list of metadatas

http://hal.in2p3.fr/in2p3-01174724
Contributor : Michel Lion <>
Submitted on : Thursday, July 9, 2015 - 3:41:33 PM
Last modification on : Thursday, February 1, 2018 - 1:35:23 AM

Identifiers

Collections

Citation

F.J. Egea Canet, V. Gonzalez, M. Tripon, M. Jastrzab, A. Triossi, et al.. A New Front-End High-Resolution Sampling Board for the New-Generation Electronics of EXOGAM2 and NEDA Detectors . IEEE Transactions on Nuclear Science, Institute of Electrical and Electronics Engineers, 2015, 62 (3), pp.1056-1062. ⟨10.1109/TNS.2015.2433021⟩. ⟨in2p3-01174724⟩