High accuracy position response calibration method for a micro-channel plate ion detector

Abstract : We have developed a position response calibration method for a micro-channel plate (MCP) detector with a delay-line anode position readout scheme. Using an {\em in situ} calibration mask, an accuracy of 8~$\mu$m and a resolution of 85~$\mu$m (FWHM) have been achieved for MeV-scale $\alpha$ particles and ions with energies of $\sim$10~keV. For high statistics experiments, this method can be directly employed with the experimental data without any dedicated calibration runs. The improved performance of the MCP detector can find applications in many fields of AMO and nuclear physics. In our case, it helps reducing systematic uncertainties in a high-precision nuclear $\beta$-decay experiment.
Type de document :
Article dans une revue
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2016, 835, pp.42-51. 〈10.1016/j.nima.2016.08.024〉
Liste complète des métadonnées

http://hal.in2p3.fr/in2p3-01323815
Contributeur : Sandrine Guesnon <>
Soumis le : mardi 31 mai 2016 - 11:16:07
Dernière modification le : mardi 5 juin 2018 - 10:14:13

Lien texte intégral

Identifiants

Collections

Citation

R. Hong, A. Leredde, Y. Bagdasarova, X. Fléchard, A. Garcia, et al.. High accuracy position response calibration method for a micro-channel plate ion detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2016, 835, pp.42-51. 〈10.1016/j.nima.2016.08.024〉. 〈in2p3-01323815〉