First experiment with the NUSTAR/FAIR Decay Total Absorption γγ-Ray Spectrometer (DTAS) at the IGISOL IV facility

Abstract : The new Decay Total Absorption Spectrometer (DTAS) has been commissioned with low energy radioactive beams at the upgraded IGISOL IV facility. The DTAS is a segmented detector composed of up to 18 NaI(Tl) crystals and it will be a key instrument in the DESPEC experiment at FAIR. In this document we report on the experimental setup and the first measurements performed with DTAS at IGISOL. The detector was characterized by means of MC simulations, and this allowed us to calculate the response function of the spectrometer and analyse the first cases of interest.
Document type :
Conference papers
Complete list of metadatas

http://hal.in2p3.fr/in2p3-01334720
Contributor : Sophie Heurteau <>
Submitted on : Tuesday, June 21, 2016 - 11:33:46 AM
Last modification on : Friday, January 11, 2019 - 11:38:39 AM

Links full text

Identifiers

Citation

V. Guadilla, A. Algora, J.L. Tain, J. Agramunt, J. Äystö, et al.. First experiment with the NUSTAR/FAIR Decay Total Absorption γγ-Ray Spectrometer (DTAS) at the IGISOL IV facility. XVIIth International Conference on Electromagnetic Isotope Separators and Related Topics (EMIS2015), May 2015, Grand Rapids, MI, United States. pp.334-337, ⟨10.1016/j.nimb.2015.12.018⟩. ⟨in2p3-01334720⟩