Abstract : The General Electronics for TPCs (GET) has been developed to equip a generation of time projection chamber detectors for nuclear physics, and may also be used for a wider range of detector types. The goal of this paper is to propose first analysis procedures to be applied on raw data samples from the GET system, in order to correct for systematic effects observed on test measurements. We also present a method to estimate the response function of the GET system channels. The response function is required in analysis where the input signal needs to be reconstructed, in terms of time distribution, from the registered output samples.
Type de document :
Article dans une revue
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2016, 840, pp.15 - 27. 〈10.1016/j.nima.2016.09.018〉
Liste complète des métadonnées

http://hal.in2p3.fr/in2p3-01402226
Contributeur : Michel Lion <>
Soumis le : jeudi 24 novembre 2016 - 14:05:55
Dernière modification le : mercredi 31 janvier 2018 - 12:29:47

Identifiants

Citation

J. Giovinazzo, T. Goigoux, S. Anvar, P. Baron, B. Blank, et al.. GET electronics samples data analysis. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, Elsevier, 2016, 840, pp.15 - 27. 〈10.1016/j.nima.2016.09.018〉. 〈in2p3-01402226〉