s'authentifier
version française rss feed
HAL : in2p3-00258718, version 1

Fiche détaillée  Récupérer au format
Nuclear Instruments and Methods in Physics Research Research: Accelerators, Spectrometers, Detectors and Associated Equipment 588 (2008) 448-456
On-line monitoring of fluence distributions and imaging of scanning ion beams
C. Pautard1, E. Balanzat2, G. Ban1, E. Batin1, B. Carniol1, J. Colin1, D. Cussol1, D. Etasse1, J.M. Fontbonne1, M. Labalme1, P. Laborie1
(2008)

In this paper we describe the design of an ion beam monitor developed to control irradiations of biological samples with ions at GANIL (Grand Accélérateur National d'Ions Lourds). This device can be used as an on-line monitor to provide transverse fluence distributions for active scanning ion irradiations. It can also be used as an interceptive beam imager to provide beam profiles or fluence depth distributions. A prototype of the monitor has been tested at GANIL with five different ion beams, having intensities from 104 to 109ions per second. Real time transverse fluence distributions have been obtained with a 1mm spatial resolution.
1 :  LPCC - Laboratoire de Physique Corpusculaire de Caen
2 :  CIMAP - UMR 6252 - Centre de recherche sur les Ions, les MAtériaux et la Photonique
Applications industrielles et médicales
Physique/Physique/Physique Médicale
Radiobiology – Radiation therapy – On-line monitor – Fluence maps – Real time ion beam imaging
Liste des fichiers attachés à ce document : 
PDF
NIM_A_Pautard_2007.pdf(81.5 KB)