Test of a Compton Telescope Prototype Based on Continuous LaBr3 Crystals and Silicon Photomultipliers
Trovato M., Barrio J., Cabello J., Callier S., Gillam J. et al
Dans Proceedings of the Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC) - 2012 Nuclear Science Symposium, Medical Imaging Conference and 19th International Workshop on Room-Temperature Semiconductor X-Ray and Gamma-Ray Detectors (2012 IEEE NSS/MIC/RTSD), Anaheim : États-Unis (2012) - http://hal.in2p3.fr/in2p3-00747232
Physique/Physique/Physique Médicale
Sciences de l'ingénieur/Micro et nanotechnologies/Microélectronique
Test of a Compton Telescope Prototype Based on Continuous LaBr3 Crystals and Silicon Photomultipliers
M. Trovato1, J. Barrio1, J. Cabello1, S. Callier2, J. Gillam1, C. De La Taille2, C. Lacasta1, M. Rafecas1, C. Solaz1, V. Stankova1, I. Torres-Espallardo1, L. Raux2, G. Llosa1
1 :  IFIC - Instituto de Fisica Corpuscular
http://ific.uv.es/
Consejo Superior de Investigaciones Científicas - CSIC (SPAIN) – Universidad de Valencia
Edificio Institutos de Investigación Apartado de Correos 22085 46071 Valencia
Espagne
2 :  LAL - Laboratoire de l'Accélérateur Linéaire
http://www.lal.in2p3.fr/
CNRS : UMR8607 – IN2P3 – Université Paris XI - Paris Sud
Centre Scientifique d'Orsay B.P. 34 91898 ORSAY Cedex
France
Omega
A Compton telescope prototype for dose monitoring in hadron therapy has been developed. The camera is based on continuous LaBr3 and LYSO crystals coupled to silicon photomultipliers for detection and readout respectively. The setup consists of a 16x18x5 mm3 LaBr3 crystal coupled to a 16x18 mm2 MPPC array as a first layer of the telescope, and a 12x12x5 mm3 LYSO crystal coupled to another 16x182 mm MPPC array as a second layer. The SPIROC1 ASIC is employed as readout electronics. The first coincidence tests have been successfully carried out and allow the reconstruction of a point-like source, in different positions, encouraging the development and the improvement of the prototype to a full imaging system.

Communications avec actes
2012
02/11/2012
Proceedings of the Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC)
internationale
3261-3264
IEEE

2012 Nuclear Science Symposium, Medical Imaging Conference and 19th International Workshop on Room-Temperature Semiconductor X-Ray and Gamma-Ray Detectors (2012 IEEE NSS/MIC/RTSD)
Poster
Anaheim
États-Unis
27/10/2012
03/11/2012

LAL/RT 12-25