Experimental results and first 22Na source image reconstruction by two prototype modules in coincidence of a liquid xenon positron emission tomograph for small animal imaging
M.L. Gallin-Martel1, Yannick Grondin2, Nicolas Gac (, http://www.lss.supelec.fr/perso/nicolas.gac)3, Y. Carcagno1, L. Gallin-Martel1, D. Grondin1, M. Marton1, J.F. Muraz1, O. Rossetto1, F. Vezzu1
1 :  LPSC - Laboratoire de Physique Subatomique et de Cosmologie
http://lpsc.in2p3.fr
CNRS : UMR5821 – IN2P3 – Université Joseph Fourier - Grenoble I – Institut Polytechnique de Grenoble - Grenoble Institute of Technology
53 avenue des Martyrs - 38026 Grenoble Cedex
France
2 :  TIMC-IMAG - Techniques de l'Ingénierie Médicale et de la Complexité - Informatique, Mathématiques et Applications, Grenoble
http://www-timc.imag.fr
CNRS : UMR5525 – Université Joseph Fourier - Grenoble I
Domaine de la Merci - 38706 La Tronche
France
3 :  Laboratoire des signaux et systèmes (L2S)
http://www.lss.supelec.fr/
UMR8506 CNRS – SUPELEC – Université Paris XI - Paris Sud
Plateau de Moulon 3 rue Joliot Curie 91192 GIF SUR YVETTE CEDEX
France
GMCAO
Division Signaux
A detector with a very specific design using liquid Xenon (LXe) in the scintillation mode is studied for Positron Emission Tomography (PET) of small animals. Two prototype modules equipped with Position Sensitive Photo Multiplier Tubes (PSPMTs) operating in the VUV range (178 nm) and at 165 K were built and studied in coincidence. This paper reports on energy, time and spatial resolution capabilities of this experimental test bench. Furthermore, these experimental results were used to perform the first image reconstruction of a 22Na source placed in the experimental setup.
Anglais
17/02/2011

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment (Nucl. Instr.& Meth. Phys. A)
Publisher Elsevier
ISSN 0168-9002 
internationale
Articles dans des revues avec comité de lecture
01/08/2012
682
66-74

Positron emission tomography (PET) – Medical imaging equipment – Liquid xenon

Liste des fichiers attachés à ce document : 
PDF
pre-print_GALLIN-MARTEL_Xenon_Liquid_PET_2012.pdf(401.2 KB)