A detection system to measure muon-induced neutrons for direct Dark Matter searches - IN2P3 - Institut national de physique nucléaire et de physique des particules Accéder directement au contenu
Article Dans Une Revue Astroparticle Physics Année : 2010

A detection system to measure muon-induced neutrons for direct Dark Matter searches

K. Eitel
  • Fonction : Auteur

Résumé

Muon-induced neutrons constitute a prominent background component in a number of low count rate experiments, namely direct searches for Dark Matter. In this work we describe a neutron detector to measure this background in an underground laboratory, the Laboratoire Souterrain de Modane. The system is based on 1 m3 of Gd-loaded scintillator and it is linked with the muon veto of the EDELWEISS-II experiment for coincident muon detection. The system was installed in autumn 2008 and passed since then a number of commissioning tests proving its full functionality. The data-taking is continuously ongoing and a count rate of the order of 1 muon-induced neutron per day has been achieved.
Fichier principal
Vignette du fichier
1006.3098.pdf (883.56 Ko) Télécharger le fichier
Origine : Fichiers produits par l'(les) auteur(s)

Dates et versions

in2p3-00493112 , version 1 (10-02-2012)
in2p3-00493112 , version 2 (11-01-2022)

Identifiants

Citer

V.Yu. Kozlov, E. Armengaud, C. Augier, Alain Benoit, L. Berge, et al.. A detection system to measure muon-induced neutrons for direct Dark Matter searches. Astroparticle Physics, 2010, 34 (2), pp.97-105. ⟨10.1016/j.astropartphys.2010.06.001⟩. ⟨in2p3-00493112v2⟩
254 Consultations
74 Téléchargements

Altmetric

Partager