A liquid Xenon Positron Emission Tomograph for small animal imaging : first experimental results of a prototype cell - IN2P3 - Institut national de physique nucléaire et de physique des particules Accéder directement au contenu
Article Dans Une Revue Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment Année : 2009

A liquid Xenon Positron Emission Tomograph for small animal imaging : first experimental results of a prototype cell

Résumé

A detector using liquid Xenon (LXe) in the scintillation mode is studied for Positron Emission Tomography (PET) of small animals. Its specific design aims at taking full advantage of the Liquid Xenon scintillation properties. This paper reports on energy, time and spatial resolution capabilities of the first LXe prototype module equipped with a Position Sensitive Photo- Multiplier tube (PSPMT) operating in the VUV range (178 nm) and at 165 K. The experimental results show that such a LXe PET configuration might be a promising solution insensitive to any parallax effect.

Dates et versions

in2p3-00331510 , version 1 (17-10-2008)

Identifiants

Citer

M.L. Gallin-Martel, L. Gallin-Martel, Y. Grondin, O. Rossetto, J. Collot, et al.. A liquid Xenon Positron Emission Tomograph for small animal imaging : first experimental results of a prototype cell. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2009, 599, pp.275-283. ⟨10.1016/j.nima.2008.11.020⟩. ⟨in2p3-00331510⟩
17 Consultations