Monte Carlo validation of a mu-SPECT imaging system on the lightweight grid CiGri - Archive ouverte HAL Access content directly
Preprints, Working Papers, ... Year : 2008

Monte Carlo validation of a mu-SPECT imaging system on the lightweight grid CiGri

(1, 2) , (3, 4) , (1) , (5) , (6) , (4) , (4) , , (2)
1
2
3
4
5
6

Abstract

Monte Carlo Simulations (MCS) are nowadays widely used in the field of nuclear medicine for system and algorithms designs. They are valuable for accurately reproducing experimental data, but at the expense of a long computing time. An efficient solution for shorter elapsed time has recently been proposed: grid computing. The aim of this work is to validate a small animal gamma camera MCS and to confirm the usefulness of grid computing for such a study. Good matches between measured and simulated data were achieved and a crunching factor up to 70 was attained on a lightweight campus grid.
Fichier principal
Vignette du fichier
FGCS-S-08-00472_0.pdf (2.48 Mo) Télécharger le fichier
Origin : Files produced by the author(s)
Loading...

Dates and versions

in2p3-00363613 , version 1 (24-02-2009)

Identifiers

  • HAL Id : in2p3-00363613 , version 1

Cite

Joe Aoun, B. Bzeznik, Vincent Breton, M. Leabad, J. Dimastromatteo, et al.. Monte Carlo validation of a mu-SPECT imaging system on the lightweight grid CiGri. 2009. ⟨in2p3-00363613⟩